DOI | Trouver le DOI : https://doi.org/10.1109/TIM.1972.4314046 |
---|
Auteur | Rechercher : Hill, Joseph J.1 |
---|
Affiliation | - Conseil national de recherches du Canada. Institut des étalons nationaux de mesure du CNRC
|
---|
Format | Texte, Article |
---|
Résumé | A theoretical investigation of the causes of error in multidecade two-staged inductive voltage dividers (IVD) has been made and from this an optimized design for the frequency band 10-400 Hz has been evaluated. A three-decade divider rated at 0.6 times frequency has been constructed and calibrated and the errors of voltage division do not exceed the design limits of 5 × 10-9 of input. |
---|
Date de publication | 1972-11-12 |
---|
Maison d’édition | IEEE |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Numéro du CNRC | NRC-INMS-4427 |
---|
Numéro NPARC | 5765535 |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | dc344c87-c6df-4abe-9cc5-6e05716b6938 |
---|
Enregistrement créé | 2009-03-29 |
---|
Enregistrement modifié | 2020-05-08 |
---|