Téléchargement | - Voir le manuscrit accepté : Mass bias fractionation laws for multi-collector ICPMS: assumptions and their experimental verification (PDF, 864 Kio)
|
---|
DOI | Trouver le DOI : https://doi.org/10.1021/ac9008435 |
---|
Auteur | Rechercher : Meija, Juris1Identifiant ORCID : https://orcid.org/0000-0002-3349-5535; Rechercher : Yang, Lu1Identifiant ORCID : https://orcid.org/0000-0002-6896-8603; Rechercher : Sturgeon, Ralph1; Rechercher : Mester, Zoltán1Identifiant ORCID : https://orcid.org/0000-0002-2377-2615 |
---|
Affiliation | - Conseil national de recherches du Canada. Institut des étalons nationaux de mesure du CNRC
|
---|
Format | Texte, Article |
---|
Résumé | Although mass bias fractionation correction is among the most studied theoretical aspects of analytical mass spectrometry, several assumptions imbedded in the conventional correction models remain largely untested. Experimental evidence is given herein highlighting parts-per-thousand deviations from the conventional mass bias correction models which can occur with use of multicollector inductively coupled plasma mass spectrometry. Furthermore, current mass spectrometric approaches are not capable of elucidating the fractionation exponent. |
---|
Date de publication | 2009-07-27 |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Numéro NPARC | 21277193 |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | abec0e70-99f4-4510-8d15-de8bfae96aba |
---|
Enregistrement créé | 2016-01-05 |
---|
Enregistrement modifié | 2024-08-12 |
---|