| DOI | Trouver le DOI : https://doi.org/10.1109/TIM.1990.1032968 |
|---|
| Auteur | Rechercher : Miljanic, P. N.1; Rechercher : Angelo, D. J.1 |
|---|
| Affiliation | - Conseil national de recherches Canada. Institut des étalons nationaux de mesure du CNRC
|
|---|
| Format | Texte, Article |
|---|
| Résumé | High precision range extenders for high voltage (600 V) and large current (100 A) have been developed as separate units to be used with a previously developed AC calibrator 111. To obtain high accuracy and stable operation an error feed-forward technique was implemented. A current-comparator-based capacitance bridge was used for the gain and phase calibration of the high voltage range extender. A precision two stage current transformer and a very stable time-division based wattconverter were u8ed for the verification of the current range extender. |
|---|
| Date de publication | 1991-04 |
|---|
| Maison d’édition | IEEE |
|---|
| Dans | |
|---|
| Langue | anglais |
|---|
| Publications évaluées par des pairs | Oui |
|---|
| Numéro du CNRC | NRC-INMS-2615 |
|---|
| Numéro NPARC | 8896507 |
|---|
| Exporter la notice | Exporter en format RIS |
|---|
| Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
|---|
| Identificateur de l’enregistrement | a76d6a58-dbc8-4d0b-8c3e-4d21454d097b |
|---|
| Enregistrement créé | 2009-04-22 |
|---|
| Enregistrement modifié | 2020-09-24 |
|---|