DOI | Trouver le DOI : https://doi.org/10.1364/AO.35.003607 |
---|
Auteur | Rechercher : Gibb, K.1; Rechercher : Decker, J. E.1; Rechercher : Boivin, L. P.1; Rechercher : Das, S. R.2; Rechercher : Buchanan, M. A.2 |
---|
Affiliation | - Conseil national de recherches du Canada. Institut des étalons nationaux de mesure du CNRC
- Conseil national de recherches du Canada. Institut des sciences des microstructures du CNRC
|
---|
Format | Texte, Article |
---|
Sujet | electrical-substitution radiometer; absolute radiometer; detector calibration; spectral responsivity; radiometric standard; thermocouple; thermopile; Seebeck coefficient; photolithography |
---|
Résumé | We describe the development of a third generation of electrical-substitution radiometers (ESR’s) at the National Research Council of Canada. The new ESR’s follow the same general design as before, but incorporate improved thermopiles and electrical heating elements. The ESR’s have a responsivity between 0.6 and 1.0 VW-1, a time constant of approximately 2.0 s, a uniformity of 0.1% over a 6-mm-diameter region, and a noise level of approximately 6 nW. Performance characteristics of the new ESR’s are discussed. It is shown that calibrations performed with these ESR’s agree with those made with the previous generation of ESR’s to better than 0.05%. |
---|
Date de publication | 1996-07 |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Numéro du CNRC | NRC-INMS-422 |
---|
Numéro NPARC | 5763987 |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | 61249839-fe90-4dda-a7c2-666551f1e5b3 |
---|
Enregistrement créé | 2009-03-29 |
---|
Enregistrement modifié | 2020-03-20 |
---|