DOI | Trouver le DOI : https://doi.org/10.1016/j.apradiso.2015.11.068 |
---|
Auteur | Rechercher : Galea, R.1; Rechercher : Gameil, K. |
---|
Affiliation | - Conseil national de recherches du Canada. Science des mesures et étalons
|
---|
Format | Texte, Article |
---|
Conférence | 20th International Conference on Radionuclide Metrology and its Applications, June 8-11 2015, Vienna, Austria |
---|
Sujet | Quality assurance; Radionuclide ionization chambers; Hospital radionuclide calibrators |
---|
Résumé | From 1986 to 2000, a fraction of the Canadian nuclear medicine community participated in a service offered by the National Research Council (NRC) of Canada to check the accuracy of administered doses of radiopharmaceuticals. The NRC renewed this service in recent years with a revalidation and reviving of the Secondary Standard Ionizing Radiation Chamber System (SSIRCS). The NRC conducted mock services for ⁹⁹ mTc, in varying geometries (syringes and serum vials), on two NRC commercial radionuclide calibrators, and at a nuclear medicine department in Canada. These tests showed the measured doses to be within 10% of the prepared standard but in some cases were sufficiently different from unity to warrant specific geometry factors to be derived. |
---|
Date de publication | 2015-11-25 |
---|
Maison d’édition | Elsevier |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Identificateur | S0969804315302931 |
---|
Numéro NPARC | 23000073 |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | 55b584c0-d7d7-4e94-ade1-9f97574baf41 |
---|
Enregistrement créé | 2016-06-01 |
---|
Enregistrement modifié | 2020-04-22 |
---|