| DOI | Trouver le DOI : https://doi.org/10.1016/S0584-8547(01)00263-4 |
|---|
| Auteur | Rechercher : Willie, S. N.1; Rechercher : Sturgeon, R.1 |
|---|
| Affiliation | - Conseil national de recherches Canada. Institut des étalons nationaux de mesure du CNRC
|
|---|
| Format | Texte, Article |
|---|
| Sujet | inductively coupled plasma; time-of-flight mass spectrometry; trace elements; rare earth elements; flow injection; on-line separation and preconcentration; seawater |
|---|
| Résumé | A method is presented for the quantitation of trace elements and rare earth elements (REE) in seawater by inductively coupled plasma orthogonal acceleration time-of-flight mass spectrometry. On-line separation and preconcentration was achieved using a timed flow-injection (FI) system incorporating a column containing iminodiacetate-based resin. The procedure was validated for selected transition elements by analysis of certified seawater reference materials NASS-5 and CASS-4. Recoveries of the rare earth elements (REE) from seawater were between 96 and 101%. Limits of detection (3s) were generally in the 20–50-pg l−1 range with nominal precision of 5–6% R.S.D. (n=5) when processing 50-ml sample aliquots of seawater. Sample throughput was 5 h−1 for the REE determinations and 15 h−1 for the transition elements of interest. |
|---|
| Date de publication | 2001-09-30 |
|---|
| Dans | |
|---|
| Langue | anglais |
|---|
| Numéro du CNRC | NRC-INMS-160 |
|---|
| Numéro NPARC | 5763129 |
|---|
| Exporter la notice | Exporter en format RIS |
|---|
| Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
|---|
| Identificateur de l’enregistrement | 2f68562f-e20b-4ea8-8a20-f8b67c85f131 |
|---|
| Enregistrement créé | 2009-03-29 |
|---|
| Enregistrement modifié | 2020-03-27 |
|---|