| DOI | Trouver le DOI : https://doi.org/10.1109/CPEM.2018.8501152 |
|---|
| Auteur | Rechercher : Granger, G.1, 2; Rechercher : Budovsky, I. |
|---|
| Affiliation | - Conseil national de recherches du Canada. Science des mesures et étalons
- Conseil national de recherches du Canada. Centre de recherche en métrologie
|
|---|
| Format | Texte, Article |
|---|
| Conférence | 2018 Conference on Precision Electromagnetic Measurements (CPEM 2018), 8-13 July 2018, Paris, France |
|---|
| Sujet | AC-DC difference; Josephson Arbitrary Waveform Synthesizer; Josephson Voltage Standard; pulsedriven; thermal transfer standard; transconductance amplifier |
|---|
| Résumé | A transconductance amplifier (TCA) with a negative admittance has been used in parallel with a commercial thermal transfer standard (TTS) to unload the voltage source, a pulse-driven AC Josephson Voltage Standard (also known as Josephson Arbitrary Waveform Synthesizer or JAWS). AC-DC difference measurements at 1 V and at frequencies from 10 Hz to 10 kHz show excellent agreement with the values obtained by a conventional method. Expanded uncertainties between 0.28 µV/V and 0.39 µV/V (k=2), depending on the frequency, have been achieved. |
|---|
| Date de publication | 2018-10-22 |
|---|
| Maison d’édition | IEEE |
|---|
| Dans | |
|---|
| Langue | anglais |
|---|
| Publications évaluées par des pairs | Oui |
|---|
| Exporter la notice | Exporter en format RIS |
|---|
| Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
|---|
| Identificateur de l’enregistrement | 1e55c830-5639-40c1-8033-aba3f190a954 |
|---|
| Enregistrement créé | 2019-03-22 |
|---|
| Enregistrement modifié | 2020-03-16 |
|---|