| Sujet | method of standard additions; uncertainty evaluation; model averaging |
|---|
| Résumé | In this study we demonstrate that the method of standard additions can provide unreliable results in chemical analysis when the linear calibration model is forced to fit the data. A well designed standard addition experiment can still yield results biased by 10% when the analyst relies only on the linear model. Recently, the Joint Committee for Guides in Metrology (GUM-6) has emphasized the need to address the uncertainty inherent to the choice of measurement models and here we show how model averaging can provide a practical way to account for model uncertainty in the method of standard additions. |
|---|
| Date de publication | 2023-05-11 |
|---|
| Maison d’édition | IOP Publishing |
|---|
| Licence | |
|---|
| Dans | |
|---|
| Langue | anglais |
|---|
| Publications évaluées par des pairs | Oui |
|---|
| Exporter la notice | Exporter en format RIS |
|---|
| Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
|---|
| Identificateur de l’enregistrement | 035380be-0529-44b8-a20c-e7ceb31f846b |
|---|
| Enregistrement créé | 2024-01-25 |
|---|
| Enregistrement modifié | 2024-01-25 |
|---|