| Download | - View accepted manuscript: Signal processing with unequally spaced data in Fourier-domain optical coherence tomography (PDF, 1.6 MiB)
|
|---|
| DOI | Resolve DOI: https://doi.org/10.1117/12.842402 |
|---|
| Author | Search for: Vergnole, Sébastien1; Search for: Lévesque, Daniel1; Search for: Sherif, Sherif S.; Search for: Lamouche, Guy1 |
|---|
| Affiliation | - National Research Council Canada. NRC Industrial Materials Institute
|
|---|
| Format | Text, Article |
|---|
| Conference | BIOS - SPIE Photonics West 2010 : San Francisco, California, USA, January 23-28, 2010 |
|---|
| Subject | Optical Coherence Tomography,; Swept Source OCT; Nonlinear optical signal processing |
|---|
| Abstract | Different algorithms for performing Fourier transforms with unequally sampled data in wavenumber space for Fourier-domain optical coherence tomography are considered. The efficiency of these algorithms is evaluated from point-spread functions obtained with a swept-source optical coherence tomography system and from computational time. Images of a 4-layer phantom processed with these different algorithms are compared. We show that convolving the data with an optimized Kaiser-Bessel window allowing a small oversampling factor before computing the fast Fourier transform provides the optimal trade-off between image quality and computational time. |
|---|
| Publication date | 2010-02-19 |
|---|
| In | |
|---|
| Series | |
|---|
| Language | English |
|---|
| Peer reviewed | Yes |
|---|
| NRC number | NRCC 52438 |
|---|
| NPARC number | 14522252 |
|---|
| Export citation | Export as RIS |
|---|
| Report a correction | Report a correction (opens in a new tab) |
|---|
| Record identifier | f108ef8a-30bd-42f1-9629-36c69253ba13 |
|---|
| Record created | 2010-03-01 |
|---|
| Record modified | 2020-04-17 |
|---|