DOI | Resolve DOI: https://doi.org/10.1109/CPEM.2016.7540556 |
---|
Author | Search for: Jung, Jae Kap; Search for: Lee, Young Seob; Search for: Kim, Kyu-Tae; Search for: So, Eddy1 |
---|
Affiliation | - National Research Council of Canada. Measurement Science and Standards
|
---|
Format | Text, Article |
---|
Conference | 2016 Conference on Precision Electromagnetic Measurements (CPEM 2016), July 10-15, 2016, Ottawa, ON, Canada |
---|
Subject | calibration; Capacitor; dissipation factor; capacitance bridge; measurement uncertainty |
---|
Abstract | An absolute method for determining the dissipation factor (DF) of a capacitor connected to a series resistor without the use of a reference capacitor with a known DF is described. The principle of the absolute measurement is based on the “elimination” of DF value of a reference capacitor by subtracting reading values from two successive measurements. This method was applied to the calibration of DF within a range of 1 × 10-5 to 1 × 10-2 at voltages up to 1 kV The verification of this method was confirmed by comparing the obtained values of DF with that measured by a known calibrated multi-frequency capacitance meter. |
---|
Publication date | 2016-07 |
---|
Publisher | IEEE |
---|